• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC PENTARAY NAV HIGH-DENSITY MAPPING ECO CATHETER; CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC PENTARAY NAV HIGH-DENSITY MAPPING ECO CATHETER; CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY Back to Search Results
Model Number D128207
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an ischemic ventricular tachycardia (isvt) ablation procedure with a pentaray nav high-density mapping eco catheter and a foreign material on usable length of the catheter was observed.It was reported that there was a plastic film on one of the splines on the catheter.The physician requested a replacement of the catheter.When the catheter was replaced, the issue was resolved, and the case continued.The issue was noticed before insertion into the body.The bwi representative does not have an image of the foreign material.The material resembled a film.The plastic box was disposed of but they still have the original box and the catheter.The foreign material was the size of the spline and covered one entire spline.The device was not used on the patient.No adverse patient consequence was reported.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number (b)(4).
 
Additional Manufacturer Narrative
The biosense webster inc.(bwi) product analysis lab received the device for evaluation on 09-sep-2022.The device evaluation was completed on 28-sep-2022.It was reported that a patient underwent an ischemic ventricular tachycardia (isvt) ablation procedure with a pentaray nav high-density mapping eco catheter and a foreign material on usable length of the catheter was observed.It was reported that there was a plastic film on one of the splines on the catheter.The physician requested a replacement of the catheter.When the catheter was replaced, the issue was resolved, and the case continued.The issue was noticed before insertion into the body.The bwi representative does not have an image of the foreign material.The material resembled a film.The plastic box was disposed of but they still have the original box and the catheter.The foreign material was the size of the spline and covered one entire spline.The device was not used on the patient.No adverse patient consequence was reported.Device evaluation details: the pentaray nav high-density mapping eco catheter was returned to biosense webster (bwi) for evaluation.Visual inspection of the returned device was performed following bwi procedures.During the visual analysis of the returned device, foreign material was not observed, this could have been lost during the decontamination process, however, this cannot be conclusively determined.A manufacturing record evaluation was performed for the finished device 30805487l number, and no internal action related to the complaint was found during the review.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTARAY NAV HIGH-DENSITY MAPPING ECO CATHETER
Common Device Name
CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-7981
MDR Report Key15344471
Report Number2029046-2022-02089
Device Sequence Number4529378
Product Code MTD
UDI-Device Identifier10846835012217
UDI-Public10846835012217
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K123837
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/10/2025
Device Model NumberD128207
Device Catalogue NumberD128207
Device Lot Number30805487L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2022
Supplement Date Received by Manufacturer09/09/2022
Initial Report FDA Received Date09/02/2022
Supplement Report FDA Received Date10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK BRAND CATHETER
Patient SexUnknown
-
-