| Model Number |
657316S |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: the locking screw could not lock onto the plate when using the fibula plate.When the screw was replaced, the locking screw was able to lock onto the plate.
|
| |
|
Additional Manufacturer Narrative
|
|
Please note the correction to d9/h3 as the device was not returned for evaluation.The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.H3 other text : device disposition unknown.
|
| |
|
Event or Problem Description
|
|
As reported: "the locking screw could not lock onto the plate when using the fibula plate.When the screw was replaced, the locking screw was able to lock onto the plate.".
|
| |
|