| Model Number |
2551H22 |
| Medical Device Problem Codes |
Fracture (1260); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that a hole was found at end of irrigation outlet of the foley catheter.Another foley catheter was inserted for the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that a hole was found at end of irrigation outlet of the foley catheter.Another foley catheter was inserted for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is confirmed, manufacturing related.No physical sample was returned, however, a photo sample as submitted.Based on the risk document a potential root cause for this event could be, "program error".No manufacturing issues or non-conformances were noted during review of the dhr that could have caused or contributed to the reported event.A label/pack review is not required as a review of the label could not have prevented the reported event.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|