Model Number A902916
Medical Device Problem Code Gel Leak (1267)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the foley kit was open and was noted that the ky jelly syringe was empty.Stated that the syringe was removed from the field, and a new ky jelly was obtained to insert the foley.There was some stickiness below the syringe in the tray which appeared to be dried out jelly.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.The device had not met specifications.The product was not used for patient treatment.Visual evaluation of the returned sample noted one opened (without original packaging), lubricating jelly syringe.Visual inspection of the sample noted no lubricating jelly residue inside the syringe.Also, noted no cap on the syringe.This does not meet specification per ip7602974, revision 8 which states "pre-filled syringes must not leak and must have cap in place".The device history record review could not be performed without a lot number.A labeling review was not performed because labelling could not have prevented the reported failure.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the foley kit was open and was noted that the ky jelly syringe was empty.Stated that the syringe was removed from the field, and a new ky jelly was obtained to insert the foley.There was some stickiness below the syringe in the tray which appeared to be dried out jelly.As per additional information received via sample form on 14sep2022, stated that while nurse was preparing to place the foley catheter it was noted that the lubricant syringe was empty and a new lubricant was obtained and used to place foley.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESTEP¿ FOLEY TRAY SYSTEM BARDEX® LUBRICATH® FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15347705
Report Number1018233-2022-06859
Device Sequence Number5037643
Product Code EZC
UDI-Device Identifier00801741074073
UDI-Public(01)00801741074073
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA902916
Device Catalogue NumberA902916
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/25/2022
Supplement Date Received by Manufacturer02/10/2023
Initial Report FDA Received Date09/02/2022
Supplement Report FDA Received Date02/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-