• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. WORKMATE¿ CLARIS¿ EP-4¿ CARDIAC STIMULATOR; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. WORKMATE¿ CLARIS¿ EP-4¿ CARDIAC STIMULATOR; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number H403023
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
During preparation for a pediatric supraventricular tachycardia procedure, with the patient on the table, the stimulator did not function as intended (data could not be sent forward) and the case was cancelled.The system was shutdown and restarted, the fiber optic cables were replaced, the connections were checked and the media converters were turned off and on which did not resolve the issue and the case was cancelled with no consequences to the patient.The procedure has been rescheduled.Another ep-4 unit was obtained from another lab which resolved the issue.
 
Additional Manufacturer Narrative
One workmate¿ claris¿ ep-4¿ cardiac stimulator was received for investigation.Testing revealed the stimulator failed to successfully execute the post (power on self-test) and communicate with the ep-4 touchscreen pc, which confirms the reported issue.Additional testing isolated the single board computer as the root cause of the communication issue.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, the root cause of the reported communication issue was isolated to abnormal functionality of the microprocessor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WORKMATE¿ CLARIS¿ EP-4¿ CARDIAC STIMULATOR
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15347805
Report Number2184149-2022-00206
Device Sequence Number12048018
Product Code DQK
UDI-Device Identifier05414734030236
UDI-Public05414734030236
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K092913
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH403023
Device Catalogue NumberH403023
Device Lot Number7894201
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2022
Supplement Date Received by Manufacturer08/26/2022
Initial Report FDA Received Date09/02/2022
Supplement Report FDA Received Date10/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/05/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-