ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE DR ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD2311-36Q
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturer reference number: 2017865-2022-22203.It was reported that patient presented in clinic for routine follow-up.Upon interrogation, it was found that patient's implantable cardioverter defibrillator longevity was overestimated by the programmer.The device was explanted prior to elective replacement indicator due to longevity overestimation.The patient was stable.
 
Additional Manufacturer Narrative
The reported event of overestimation was confirmed.As received, the device was above the elective replacement indicator (eri).Longevity analysis found the battery depletion to be normal.The device functioned normally during analysis with no anomalies found.The longevity overestimation was determined to be due to an estimation inaccuracy in the merlin programmer remaining longevity calculation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE DR ICD
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15348246
Report Number2017865-2022-22202
Device Sequence Number5134380
Product Code NVZ
UDI-Device Identifier05414734505994
UDI-Public05414734505994
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2014
Device Model NumberCD2311-36Q
Device Catalogue NumberCD2311-36Q
Device Lot Number3755672
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2022
Supplement Date Received by Manufacturer10/07/2022
10/14/2022
Initial Report FDA Received Date09/02/2022
Supplement Report FDA Received Date10/12/2022
10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DURATA
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexFemale
Patient Weight68 KG
Patient EthnicityNon Hispanic
-
-