• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, LAMP¿ 90, 63 CM LENGTH, 8.5F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, LAMP¿ 90, 63 CM LENGTH, 8.5F; INTRODUCER, CATHETER Back to Search Results
Model Number 407358
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One 8.5f swartz braided introducer sheath was received for evaluation.Functional testing confirmed an air leak and fluid leak in the hemostasis valve.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seals and subsequent leak remains unknown.
 
Event or Problem Description
This report it to advise of a leak in the sheath during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, LAMP¿ 90, 63 CM LENGTH, 8.5F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15348611
Report Number3005334138-2022-00500
Device Sequence Number4529658
Product Code DYB
UDI-Device Identifier05414734205535
UDI-Public05414734205535
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K052644
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model Number407358
Device Catalogue Number407358
Device Lot Number7788307
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2022
Initial Report FDA Received Date09/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-