No product sample was received; therefore, visual, and functional testing could not be performed.A review of the production floor was conducted.Samples of tubing sets were taken and visually inspected using a black light to verify that the solvent was in each join and the amount was adequate; no discrepancies were detected.Samples were taken and visually inspected with the camera vision system to verify that the skin adhesive was properly assembled; no discrepancies were detected.A device history record (dhr) review was performed and found no discrepancies or abnormalities relevant to the complaint.No root cause could be determined as the complaint could not be confirmed.No information has been received to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.
|