| Model Number |
TRULIANT TIB IMP CRC INSERT SZ 3.5, 11MM |
| Medical Device Problem Codes |
Unstable (1667); Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Synovitis (2094); Joint Laxity (4526)
|
| Date of Event |
08/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, the patient had an initial tka on (b)(6) 2020.The patient showed up to the clinic with pain and synovitis.The patient was revised on (b)(6) 2022.Laxity was noted.There was some pitting and surface changes noted on the insert.The patient was stable and had good rom.A new insert was implanted.The patient left in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending investigation.Concomitant medical products: 02-020-13-0235 - truliant cr cem fem cr cem left sz 3.5 sn: (b)(4), 02-022-45-3525 - truliant tib fit tray cem sz 3.5f / 2.5t sn: (b)(4), 200-02-35 - three peg patella 35mm sn: (b)(4), 201-78-15 - holding pin mini sharp point 4 pk sn: (b)(4), 521-78-23 - threaded pin size 2.3 collared 2pk sn: (b)(4), 521-78-32 - threaded pin size 3.0 collarless 2pk sn: (b)(4), 521-78-32 - threaded pin size 3.0 collarless 2pk sn: (b)(4), 521-78-36 - threaded pin size 5.1 collarless 2pk sn: (b)(4), a10012 - gps implant kit v2 sn: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The patient has bilateral knee replacements, this suggests joint maladies.The cause of the subsequent revision cannot be conclusively determined; however, it is most likely related to the patient¿s condition.H7: z-0023-2022.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: a, b, c, d, g.The revision reported may have been due to prosthesis wear, and/or instability.However, the reported instability could not be confirmed.Additionally, potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and relevant clinical information was not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|