MEDIVATORS, INC. ADVANTAGE PLUS ENDOSCOPE REPROCESSING SYSTEM Back to Search Results
Model Number EGA-500
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A steris service technician arrived onsite to inspect the unit and found that the connectors and o-ring on one of the water filters were broken.The technician replaced the connectors and o-ring on the filter, tested the unit, confirmed it to be operating according to specification, and returned it to service.No additional issues have been reported.
 
Event or Problem Description
The user facility reported that their advantage plus endoscope reprocessing system was leaking water onto the floor.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTAGE PLUS ENDOSCOPE REPROCESSING SYSTEM
Common Device Name
ENDOSCOPE REPROCESSING SYSTEM
Manufacturer (Section D)
MEDIVATORS, INC.
14605 28th avenue n
minneapolis MN 55447
Manufacturer (Section G)
MEDIVATORS, INC.
14605 28th avenue n
minneapolis MN 55447
Manufacturer Contact
daniel davy
14605 28th ave n
minneapolis, MN 55447
4403927453
MDR Report Key15349270
Report Number2150060-2022-00035
Device Sequence Number5227346
Product Code FEB
UDI-Device Identifier00677964063664
UDI-Public00677964063664
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEGA-500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2022
Initial Report FDA Received Date09/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/03/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-