| Model Number |
MIDDLE SEGMENT TRIAL 25 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During surgical use, upon removal of the trial segment from the stem in situ, the 4 "legs" of the trial broke.The pieces were retrieved and accounted for.No surgical delay/prolongation.The patient was last known to be in stable condition following the event.No other patient information/medical history reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending investigation.Pending device return.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: as determined, the broken instrument reported was likely the result of not considering specific performance requirements to capture user needs and design inputs (material durability during reprocessing/sterilization as well as force applied over lifetime of device), which led to insufficient testing to demonstrate continued
performance over time and resulted in fracture during surgical use.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections: b1, b2, h1.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment of the reported event, it was determined that the event did not involve a death or serious injury nor would be likely to cause or contribute to a death or serious injury if the reported malfunction were to recur.As a result, this complaint event is no longer considered reportable by exactech and this report may be disregarded.
|
| |
|