MEDTRONIC EUROPE SARL COBALT¿ HF QUAD CRT-D MRI SURESCAN¿; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTPB2QQ |
Device Problems
Device Sensing Problem (2917); Reset Problem (3019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/06/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) exhibited a full reset, and the device settings were automatically adjusted back to nominal settings.It was also noted that inappropriate internal programming of a wavelet template parameter occurred.When the patient was brought to the clinic to have the device programmed back to the original settings an x-ray was performed and it was found that all the leads had pulled back.The right atrial (ra) lead exhibited high thresholds and was not capturing at high outputs.The x-ray confirmed the lead had dislodged.A revision was attempted but too difficult to maneuver so the ra lead was explanted and replaced.The right ventricular (rv) lead trends were stable but due to having pulled back slightly it was repositioned.The left ventricular (lv) lead also exhibited high thresholds and it was suspected to have slightly pulled back.The lead was attempted to be repositioned.When this attempt to reposition failed, two different lv lead were attempted but could not be placed due to patient anatomy.Due to the inability to place a new lv lead the device was downgraded to an implantable cardioverter defibrillator (icd).The device was originally reprogrammed and later explanted and replaced due to downgrade.The rv lead remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|