Model Number 362725
Medical Device Problem Code Short Fill (1575)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported when using the bd vacutainer® est z (no additive) plus blood collection tube there was under-fill or low draw of a tube with blood.The following information was provided by the initial reporter.The customer stated: "we have been alerted by several departments of the hospital centre of a lack of vacuum for purge tubes (ref (b)(4) of batch 2056870.".
 
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd vacutainer® est z (no additive) plus blood collection tube there was under-fill or low draw of a tube with blood.The following information was provided by the initial reporter.The customer stated: "we have been alerted by several departments of the hospital centre of a lack of vacuum for purge tubes (ref 362725) of batch 2056870.".
 
Additional Manufacturer Narrative
H.6.Investigation summary: bd had not received samples or photos for investigation.Therefore, 20 retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to underfill as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode underfill.Bd was not able to identify a root cause for the indicated failure mode.
 
Event or Problem Description
It was reported when using the bd vacutainer® est z (no additive) plus blood collection tube there was under-fill or low draw of a tube with blood.The following information was provided by the initial reporter.The customer stated: "we have been alerted by several departments of the hospital centre of a lack of vacuum for purge tubes (ref 362725) of batch 2056870.".
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for eval: yes.D9: returned to manufacturer on: 2022-09-20.H.6.Investigation summary: bd received 11 samples from the customer in support of this complaint.The 11 customer samples along 20 retained samples from the bd inventory were functionally tested and the issue of underfill was not observed.Bd was unable to confirm customers indicated failure mode based on the returned and retained samples test results.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.Information will be captured on trend reports and monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® EST Z (NO ADDITIVE) PLUS BLOOD COLLECTION TUBE
Common Device Name
SPECIMEN TRANSPORT AND STORAGE CONTAINER
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15351387
Report Number9617032-2022-00824
Device Sequence Number5863249
Product Code FMH
UDI-Device Identifier50382903627252
UDI-Public50382903627252
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2023
Device Model Number362725
Device Catalogue Number362725
Device Lot Number2056870
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2022
Supplement Date Received by Manufacturer09/09/2022
09/22/2022
Initial Report FDA Received Date09/02/2022
Supplement Report FDA Received Date09/20/2022
10/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-