As reported, during an unspecified procedure, a cxi support catheter separated.Access was obtained in the femoral artery and an ipsilateral approach was used to treat a severely calcified lesion in the iliac artery.A 6 french sheath was used for access.The complaint device would not advance through the lesion and was therefore removed.After the catheter was removed from the patient, it was noted to be kinked and separated approximately fifteen centimeters from the distal end of the device.The separated fragment of the device remained in the iliac artery.A balloon catheter was then advanced to the distal side of the separated cxi fragment and the balloon was inflated, allowing the separated fragment to be pulled with the inflated balloon and removed from the patient.Another manufacturer's device was used to complete the procedure.There have been no adverse effects to the patient.
|
|
Blank fields on this form indicate the information is unknown or unavailable.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event summary: as reported, during an unspecified procedure, a cxi support catheter separated.Access was obtained in the femoral artery and an ipsilateral approach was used to treat a severely calcified lesion in the iliac artery.A 6 french sheath was used for access.The complaint device would not advance through the lesion and was therefore removed.After the catheter was removed from the patient, it was noted to be kinked and separated approximately fifteen centimeters from the distal end of the device.The separated fragment of the device remained in the iliac artery.A balloon catheter was then advanced to the distal side of the separated cxi fragment, and the balloon was inflated, allowing the separated fragment to be pulled with the inflated balloon and removed from the patient.Another manufacturer's device was used to complete the procedure.There have been no adverse effects to the patient.Investigation ¿ evaluation.Reviews of the complaint history, device history record, instructions for use, manufacturing instructions and quality control procedures were conducted during the investigation.The device was not returned for investigation.A document-based investigation evaluation was performed.No related non-conformances were recorded.No additional complaints were received for this product lot.The device history record review provides objective evidence that the device was manufactured to specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was functionally inspected by quality control and no notable gaps in production or processing controls were noted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.The device is provided with instructions for use which caution, "the catheter should not be advanced in an area of resistance unless the source of the resistance is identified under fluoroscopy and appropriate steps are taken to reduce or remove the obstruction.¿ based on the available information, cook has concluded that the patient¿s anatomy was the cause of the device failure in this incident.The complaint device was unable to pass through the severely calcified lesion, a separation occurred during removal, and a kink was found after.The complaint was confirmed based on customer testimony.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per a review of risk documentation, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|