| Model Number |
4003 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the device provided inaccurate spo2 readings.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the device has not been returned to the investigation facility to allow for an analysis to be performed.If the product is received for evaluation or new information is obtained, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported the device provided inaccurate spo2 readings.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing. during simulation testing, the sensor passed manual and preset conditions and provided accurate measurements.The sensor was determined to be functioning as designed. .
|
| |
|