Model Number 70914-70
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product has been returned, extended investigation is pending at this time.A follow- up report will be submitted once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of 397 mg/dl, 199 mg/dl, 97 mg/dl, 404 mg/dl, and 209 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
 
Additional Manufacturer Narrative
The reported meter and test strips have been returned and investigated.Performed visual inspection on returned meter and returned strips.Meter turned on with button depression and test strip insertion.Control solution testing was performed and all results were within range specification.No malfunction or product deficiency was identified.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The useful life of freestyle freedom lite meter is 5 years.As the manufacturing date of this product is 2010, it has been in distribution beyond its useful life at the time of the complaint.Since the product exceeded its useful life, it is determined to have met specification when the product was released and through its lifespan.No further investigation activities are being performed at this time.Issue is not confirmed.Section d4 (serial number) and h4 (device mfg date) were updated based on returned product download.Section d4 (serial number) was updated from (b)(6) to (b)(6).Section h4 (device mfg date) was updated from 6/8/2013 to 6/8/2010.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of 397 mg/dl, 199 mg/dl, 97 mg/dl, 404 mg/dl, and 209 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE FREEDOM LITE
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key15352610
Report Number2954323-2022-32331
Device Sequence Number17643012
Product Code NBW
UDI-Device Identifier00699073708229
UDI-Public00699073708229
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K092602
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/31/2023
Device Model Number70914-70
Device Catalogue Number70822
Device Lot Number1171101
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer10/03/2022
Initial Report FDA Received Date09/02/2022
Supplement Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/08/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
Patient Weight89 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-