| Model Number |
21-4455-24 |
| Medical Device Problem Codes |
Difficult to Flush (1251); Fracture (1260); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient was taken to the operating room for replacement of the device, secondary to the device clotting off, would not flush with extravasation of fluid around the reservoir.The patient had received neoadjuvant therapy since placement of the catheter.Upon removal of the catheter, it was noticed to have a fracture in the midportion of the catheter which would have been present from manufacture date.Serious injury due to failure of implant was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.
|
| |
|