The product was not returned to abbott vascular for analysis.A review of the electronic lot history record (elhr), corrective action tracking system for the web (catsweb) database review and similar incident/complaint query for this product was not performed since the lot number was not reported.The investigation determined the reported difficulties and subsequent treatment appear to be related to the operational context of the procedure.In this case, it is likely that the guide wire interacted with the moderately calcified and moderately tortuous lesion during removal, becoming jailed, resulting in the reported difficult to remove.Manipulation of the device during its interaction with the challenging anatomy likely resulted in the reported material separation and stretched coils.Multiple wires were advanced and wrapped over the separated guide wire tip to remove it from the anatomy.Additionally, the reported removal of foreign body was due to a separated portion of the guide wire within the anatomy.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|