| Model Number |
35W080 |
| Medical Device Problem Codes |
Delivered as Unsterile Product (1421); Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.Two (2) product samples were received for evaluation.Visual inspection of the pouch seal did reveal openings.The pouch was cut open at the vendor's seal and positioned on the burst tester.Under functional testing of the seal, the seal did exceed the minimum burst test.The root cause of the reported issue could not be confirmed.The investigation does not identify a manufacturing issue with the returned samples.An internal non-conformance report (ncr) has been opened to address the multiple complaints received by a single complainant.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the sterility of the device was compromised, as the bottom of the package was not sealed completely and the tube sticks out of the end.No patient injury was reported.
|
| |
|