| Model Number |
35W080 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the sterility of the device was compromised, as the bottom of the package was not sealed completely and the tube sticks out of the end.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Seven samples were received for evaluation.Visual testing was performed.Visual inspection found the pouch seals revealed an opening on the bottom seal on each pouch.The root cause of the issue could not be determined.An internal non-conformance report (ncr) has been opened to address the multiple complaints received by a single complainant.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional investigative action will be taken accordingly.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|