• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JUNI OX FB FEM SZ 5 RM LL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JUNI OX FB FEM SZ 5 RM LL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71422355
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Joint Dislocation (2374); Joint Laxity (4526)
Date of Event 10/08/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that after a left tka performed on (b)(6) 2013 due to osteoarthritis, the patient experienced dislocation, instability and progressive arthritis that was addressed via revision surgery on (b)(6) 2017.All components were explanted.This information was provided by the national joint registry of the united kingdom, as part of a retrospective data collection of patients who underwent a revision surgery of journey uni prothesis.As such, no further information will be available.
 
Additional Manufacturer Narrative
Section d10 was updated.Section h10: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a left revision was performed approximately 4 years post unicompartmental knee arthroplasty due to dislocation, instability, and progressive arthritis.As of the date of this medical investigation, the requested clinical documentation has not been received as the information was provided by the national joint registry of the united kingdom as part of a retrospective data collection, and reportedly, no further information will be available.Without the requested medical documentation, definitive clinical factors which could have contributed to the reported event could not be concluded.The patient outcome beyond that which was reported could not be determined; therefore, no further medical assessment can be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that loosening as a result from trauma, improper implant selection, improper implant positioning, improper fixation, and/or migration of the components.Muscle and fibrous tissue laxity can also contribute to these conditions, this has been identified as possible adverse effects.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to fit/sizing issue, joint laxity, traumatic injury, limited range of motion, patient anatomy, abnormal loading of limb or patient condition.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUNI OX FB FEM SZ 5 RM LL
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15354159
Report Number1020279-2022-03947
Device Sequence Number6057782
Product Code HSX
UDI-Device Identifier03596010627414
UDI-Public03596010627414
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K073175
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/23/2023
Device Model Number71422355
Device Catalogue Number71422355
Device Lot Number13EM15863
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer09/08/2022
Initial Report FDA Received Date09/05/2022
Supplement Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/23/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71422265 JOURNEY UNI TIBINRT S5-6RM/LL8MM; 71422436 JRNY UNI TIBIAL BASE RM/LL SZ 6; JOURNEY UNI TIBINRT S5-6RM/LL8MM,, LOT 12BM08025; JRNY UNI TIBIAL BASE RM/LL SZ 6, LOT 11LM03787
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age66 YR
Patient SexMale
-
-