• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRACLIP DUAL TRIGGER; BIOPSY INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRACLIP DUAL TRIGGER; BIOPSY INSTRUMENT Back to Search Results
Model Number 864017D
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.
 
Event or Problem Description
It was reported that prior to a breast tissue marker placement procedure, the marker components were allegedly disassembled.There was no patient contact.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported mislabeled product could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that prior to a breast tissue marker placement procedure, the marker components were allegedly mislabeled.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRACLIP DUAL TRIGGER
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15354352
Report Number2020394-2022-00710
Device Sequence Number13890809
Product Code NEU
UDI-Device Identifier00801741032257
UDI-Public(01)00801741032257
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K090547
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number864017D
Device Catalogue Number864017D
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/16/2022
Supplement Date Received by Manufacturer10/12/2022
Initial Report FDA Received Date09/05/2022
Supplement Report FDA Received Date10/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-