| Model Number |
DL31F-VO |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
Air Embolism (1697)
|
| Date of Event |
05/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Pt supported on vv ecmo with a dlca since 1/31.The team reported an event today 5/19 where the cannula separated at the venous y.They attempted to tie band the connection but were unable to secure and so the patient was recannulated to va ecmo.Patient welfare information - off ecmo - received a bilateral lung transplant.
|
| |
|
Additional Manufacturer Narrative
|
|
Moved warning to first page of the ifu: do not allow alcohol, alcohol-based liquids, ethers, halogenated liquids or corrosive solvents to come into contact with the device.This would cause damage such as to compromise the integrity and functionality of the device.
|
| |
|