AOMORI OLYMPUS CO., LTD. SINGLE USE REPOSITIONABLE CLIP; SINGLE USE REPOSITIONAL CLIP Back to Search Results
Model Number HX-202UR
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device was not sent to olympus for evaluation and the lot number was not provided, therefore, the manufacturing date could not be determined.A review of the device history record could not be conducted as the serial number was not provided.A review of dhr for the year prior to the date of the occurrence were reviewed and no deviations were found that could have caused or contributed to the reported event.Based on the results of the legal manufacturer's investigation, it is likely the clip remained in the scope because it was not properly released and reprocessing was insufficient to remove it.However, a definitive root cause of the reported issue could not be identified.Olympus will continue to monitor field performance for this device.Investigation activities have been opened to manage the actions related to this issue and any required mdr reporting.
 
Event or Problem Description
The customer reported that, during reprocessing of the endoscope, a clip fell out of the endoscope.The endoscope had previously been used in a polypectomy procedure during which a clip fell off and did not attach.A second clip was used to complete that procedure and the endoscope was reprocessed and placed back in service.The next case in which the endoscope was used did not require any clips.After that procedure, the endoscope was reprocessed and a clip fell out during reprocessing.There was no effect on the patient due to the event.The endoscope is being reported on the medwatch with patient identifier (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE REPOSITIONABLE CLIP
Common Device Name
SINGLE USE REPOSITIONAL CLIP
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer Contact
masaharu hirose
2-248-1 okkonoki
kuroishi-shi, aomori 036-0-357
JA   036-0357
426422891
MDR Report Key15355496
Report Number9614641-2022-00269
Device Sequence Number1727544
Product Code PKL
UDI-Device Identifier04953170345180
UDI-Public04953170345180
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K123601
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHX-202UR
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2022
Initial Report FDA Received Date09/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-