Model Number ASC4730-01
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during an arthroscopy, it was noticed that the cut was not taking place correctly and the tip of the multivac wand was found broken, the metallic part of the tip was found inside the patient.The small piece of metal was retrieved form the patient by being aspirated.Surgery was resumed after a non-significant surgical delay using a back-up device instead.No further complications were reported.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states an undated photo of the device was provided for review and confirms the reported breakage.Per case details, the broken tip was retrieved from the patient.An intraoperative photo was also provided for review which demonstrates the multivac wand being used at the hip but does not indicate root cause of the broken tip.Conclusion: the image provided was reviewed and does not contribute to a root cause of the breakage, no clinical factors were found which would have contributed to the reported event.All documents and images provided as of this date have been reviewed and considered and unless noted do not contribute to the clinical/medical investigation.No further impact to the patient is anticipated as it was noted the broken piece was retrieved.Should additional information become available this issue can be re-elevated.No further clinical assessment is warranted at this time.Factors that could have contributed to the reported event include: hitting the device tip against a hard surface.Using the device as a lever to enlarge surgical site.Mechanical displacement of tissue through applied force.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIVAC 50 XL ICW
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15355759
Report Number3006524618-2022-00390
Device Sequence Number16576043
Product Code GEI
UDI-Device Identifier00817470006261
UDI-Public00817470006261
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K033584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/14/2024
Device Model NumberASC4730-01
Device Catalogue NumberASC4730-01
Device Lot Number2085814
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2022
Supplement Date Received by Manufacturer08/09/2023
Initial Report FDA Received Date09/05/2022
Supplement Report FDA Received Date08/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-