Model Number 24686
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the mildly tortuous and mildly calcified venous anastomosis.A 8.0 x80, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during initial inflation at 18 atmospheres, the balloon ruptured.The device was aspirated and removed.The procedure was completed successfully with another 8x80 gladiator device.There were no patient complications reported.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the mildly tortuous and mildly calcified venous anastomosis.A 8.0 x80, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during initial inflation at 18 atmospheres, the balloon ruptured.The device was aspirated and removed.The procedure was completed successfully with another 8x80 gladiator device.There were no patient complications reported.It was further reported that the balloon was inflated for 30 seconds.
 
Additional Manufacturer Narrative
D4: lot number: corrected from 28872901 to 28764143 d4: expiration date: corrected from 02/14/2025 to 01/27/2025 h4: device manufacture date: corrected from 02/15/2022 to 01/28/2022.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the mildly tortuous and mildly calcified venous anastomosis.A 8.0 x80, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during initial inflation at 18 atmospheres, the balloon ruptured.The device was aspirated and removed.The procedure was completed successfully with another 8x80 gladiator device.There were no patient complications reported.It was further reported that the balloon was inflated for 30 seconds.
 
Additional Manufacturer Narrative
D4: lot number: corrected from 28872901 to 28764143.D4: expiration date: corrected from 02/14/2025 to 01/27/2025.H4: device manufacture date: corrected from 02/15/2022 to 01/28/2022.Device evaluated by mfr.: a gladiator device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The rated burst pressure for this device is 20 atm atmospheres as per gladiator elite specification.The returned device was attached to an encore inflation unit and positive pressure was applied and liquid was observed to be leaking from a balloon pinhole located approximately 25mm distally of the proximal marker band.An examination of the balloon material and proximal marker band identified no issues which could potentially have contributed to this complaint.A visual examination observed no damage to the tip of the device.A visual and tactile examination of the shaft of the device found no issues.A visual examination found no issues or damage to the markerbands of the device.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the mildly tortuous and mildly calcified venous anastomosis.A 8.0 x80, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during initial inflation at 18 atmospheres, the balloon ruptured.The device was aspirated and removed.The procedure was completed successfully with another 8x80 gladiator device.There were no patient complications reported.It was further reported that the balloon was inflated for 30 seconds.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15357706
Report Number2124215-2022-34084
Device Sequence Number1622752
Product Code LIT
UDI-Device Identifier08714729809913
UDI-Public08714729809913
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/27/2025
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0028764143
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/30/2022
Supplement Date Received by Manufacturer09/02/2022
09/21/2022
10/06/2022
Initial Report FDA Received Date09/05/2022
Supplement Report FDA Received Date09/07/2022
09/27/2022
10/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexFemale
Patient RaceBlack Or African American
-
-