• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JUNI OX FB FEM SZ 7 LM RL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JUNI OX FB FEM SZ 7 LM RL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71422347
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 10/29/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after a left tka performed on (b)(6) 2019 due to osteoarthritis, the patient experienced infection.This was addressed via revision surgery on (b)(6) 2020 were all components where exchange.This information was provided by the national joint registry of the (b)(6), as part of a retrospective data collection of patients who underwent a revision surgery of journey uni prothesis.As such, no further information will be available.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
Section h10: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a left revision was performed approximately 11 months post uka due to infection.As of the date of this medical investigation, the requested clinical documentation has not been received as the information was provided by the national joint registry of the united kingdom as part of a retrospective data collection, and reportedly, no further information will be available.Without the requested medical documentation, the source of the reported infection could not be concluded.The patient impact beyond that which was reported could not be determined.No further medical assessment is warranted at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that acute post-surgical wound infection has been identified in possible adverse effects.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the sterilization records revealed the batch was sterilized within normal parameters.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include loss of sterility during procedure, patient reaction, and post-operative healing issue.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: case(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUNI OX FB FEM SZ 7 LM RL
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15358796
Report Number1020279-2022-03975
Device Sequence Number5889062
Product Code HSX
UDI-Device Identifier03596010627360
UDI-Public03596010627360
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K073175
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Other,Foreign,Study,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71422347
Device Catalogue Number71422347
Device Lot Number18HM20455
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer10/06/2022
Initial Report FDA Received Date09/05/2022
Supplement Report FDA Received Date10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/28/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
JOURNEY UNI TIBINRT S5-6LM/RL8MM, LOT#:18AM22872; JRNY UNI TIBIAL BASE LM/RL SZ 5, LOT#:18CM02598
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age62 YR
Patient SexMale
-
-