• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD MICROTAINER® CONTACT-ACTIVATED LANCET; MANUAL SURGICAL INSTRUMENT FOR GENERAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD MICROTAINER® CONTACT-ACTIVATED LANCET; MANUAL SURGICAL INSTRUMENT FOR GENERAL USE Back to Search Results
Model Number 366594
Medical Device Problem Codes Retraction Problem (1536); Defective Component (2292); Improper Flow or Infusion (2954)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Oem manufacturer: the manufacturing location for this product is nipro.This site is an oem manufacturing site.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd microtainer® contact-activated lancet lancet did not activate, the needle was not recovered.This event occurred 200 times.The following information was provided by the initial reporter.The customer stated: "during using the lancet, it was found that the lancet did not activate, the needle was not recovery, and occurred insufficient blood flow issue.Patients and users were not affected.".
 
Event or Problem Description
It was reported when using the bd microtainer® contact-activated lancet lancet did not activate, the needle was not recovered.This event occurred 200 times.The following information was provided by the initial reporter.The customer stated: "during using the lancet, it was found that the lancet did not activate, the needle was not recovery, and occurred insufficient blood flow issue.Patients and users were not affected.".
 
Additional Manufacturer Narrative
H.6.Investigation summary: material # 366594.Lot/batch # b7b82d4.Bd received 10 samples for investigation.The samples were evaluated by visual examination and functional testing and the indicated failure modes for does not activate/does not recover/insufficient blood flow with the incident lot was not observed.Additionally, retention samples from bd inventory were evaluated by visual examination and functional testing and no issues were observed relating to does not activate/does not recover/insufficient blood flow as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure modes.Bd was not able to identify a root cause for the indicated failure modes.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICROTAINER® CONTACT-ACTIVATED LANCET
Common Device Name
MANUAL SURGICAL INSTRUMENT FOR GENERAL USE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15358920
Report Number2243072-2022-01470
Device Sequence Number12068989
Product Code FMK
UDI-Device Identifier50382903665940
UDI-Public50382903665940
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/31/2026
Device Model Number366594
Device Catalogue Number366594
Device Lot NumberB7B82D4
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer09/12/2022
Initial Report FDA Received Date09/05/2022
Supplement Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-