Model Number 2975-00-800
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the reclaim insertion handle would not fully seat onto reclaims.The case was completed successfully.There was a surgical delay of less than two minutes.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was returned for analysis.Visual examination of the device confirmed the complaint.The initial threads of the distal tip are stripped, the damaged condition prevents the instrument from properly assemble with the mating devices.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM DISTAL STEM INSERTER
Common Device Name
RECLAIM INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15360308
Report Number1818910-2022-17144
Device Sequence Number14832309
Product Code HWR
UDI-Device Identifier10603295156161
UDI-Public10603295156161
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2975-00-800
Device Catalogue Number297500800
Device Lot NumberA0411
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2022
Supplement Date Received by Manufacturer10/04/2022
Initial Report FDA Received Date09/06/2022
Supplement Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/15/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-