• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 7" (18 CM) APPX 0.72 ML, PRESSURE INFUSION (400PSIG) EXT SET W/MICROCLAVE®, PURP; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 7" (18 CM) APPX 0.72 ML, PRESSURE INFUSION (400PSIG) EXT SET W/MICROCLAVE®, PURP; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number B3361
Medical Device Problem Codes Material Discolored (1170); Particulates (1451)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is available for evaluation, it has not been received.Initial reporter name and address: initial reporter phone has an extension number of (b)(6).
 
Event or Problem Description
The event involved a 7" (18 cm) appx 0.7 ml, pressure infusion (400psig) ext set w/microclave¿, purple clamp, rotating luer where it was reported gunk in the end of the connector when opening the tubing package.It was also reported that the tubing was discolored.There was no patient involvement, no adverse event, no harm reported and no delay in therapy.
 
Additional Manufacturer Narrative
The device was not received for evaluation.The complaint of black material on the luer of the male luer can be confirmed.Received one photo showing the male luer of the b3361 with black material observed on the tip of the luer.The material is not in the fluid path.As well the tubing appeared to have a brownish discoloration.The tubing was still translucent.The discoloration is consistent with the discoloration from sterilization.Slight discoloration of tubing is a normal result of the sterilization and doesn't affect the safety or sterility of the set.The probable cause of the black material is due to burnt material embedded in the wall of the male luer during the molding process in manufacturing.The device history record (dhr) was reviewed and no non conformities were found that would have led to the reported complaint.Additional information: d9 - device available for evaluation and d9 - date returned to mfg null.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
7" (18 CM) APPX 0.72 ML, PRESSURE INFUSION (400PSIG) EXT SET W/MICROCLAVE®, PURP
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
michael visocnik
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key15362313
Report Number9617594-2022-00256
Device Sequence Number14918743
Product Code FPA
UDI-Device Identifier00887709008410
UDI-Public(01)00887709008410(17)231201(10)3923457
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K941190
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2023
Device Model NumberB3361
Device Catalogue NumberB3361
Device Lot Number3923457
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2022
Supplement Date Received by Manufacturer09/23/2022
Initial Report FDA Received Date09/06/2022
Supplement Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-