• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. FUSED LIGHT CABLE, 5.0 MM X 274CM (9.0'); LIGHT, SURGICAL, FIBEROPTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. FUSED LIGHT CABLE, 5.0 MM X 274CM (9.0'); LIGHT, SURGICAL, FIBEROPTIC Back to Search Results
Model Number FUSED LIGHT CABLE, 5.0 MM X 274CM (9.0')
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On (b)(6) 2022, it was reported by a sales representative via sems that (2) ar-3240-5027 fused light cables is smoking when plugged in, this occurred during an unknown time and was not used.
 
Additional Manufacturer Narrative
Complaint confirmed.One unpackaged ar-3240-5027, serial/batch number (b)(6), was received for investigation.Functional testing was not conducted due to connecting cable damage.Visual evaluation found that the wolf female end tip adapter was broken off before to receive for investigation.The missing part was not returned for investigation.The root cause of the reported failure mode is undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSED LIGHT CABLE, 5.0 MM X 274CM (9.0')
Common Device Name
LIGHT, SURGICAL, FIBEROPTIC
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15362657
Report Number1220246-2022-05460
Device Sequence Number14924692
Product Code FST
UDI-Device Identifier00888867029408
UDI-Public00888867029408
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K901035
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFUSED LIGHT CABLE, 5.0 MM X 274CM (9.0')
Device Catalogue NumberAR-3240-5027
Device Lot NumberWO175826
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2022
Supplement Date Received by Manufacturer08/15/2022
Initial Report FDA Received Date09/06/2022
Supplement Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/18/2021
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-