• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. CHONDRAL PICK, ANKLE ARTHROSCOPY, 60 DEG; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. CHONDRAL PICK, ANKLE ARTHROSCOPY, 60 DEG; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number CHONDRAL PICK, ANKLE ARTHROSCOPY, 60 DEG
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On (b)(6) 2022, it was reported by a sales representative via sems that a ar-8655-06 chondral pick broke, and a ar-8655-05 chondral pick was found to be bent.This occurred on (b)(6) 2022 during an unknown case when the ar-8655-06 broke and the ar-8655-05 was found to be bent.There was no patient effect reported.No additional information was provided.Additional information requested.
 
Additional Manufacturer Narrative
Complaint confirmed.Upon visual inspection, it was noted that the distal tip of the device had broken off.The fragment was not returned for analysis.It was observed that the shaft of the pick was bent, indicating that the device had been subjected to prying/leveraging forces.As such, this event is consistent with user-applied mechanical forces to the device during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHONDRAL PICK, ANKLE ARTHROSCOPY, 60 DEG
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15362688
Report Number1220246-2022-05459
Device Sequence Number5467516
Product Code LXH
UDI-Device Identifier00888867044296
UDI-Public00888867044296
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCHONDRAL PICK, ANKLE ARTHROSCOPY, 60 DEG
Device Catalogue NumberAR-8655-06
Device Lot Number1392037
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2022
Supplement Date Received by Manufacturer08/16/2022
Initial Report FDA Received Date09/06/2022
Supplement Report FDA Received Date11/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/15/2020
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-