C.R. BARD, INC. (COVINGTON) -1018233 BARD® RUBBER UTILITY CATHETER; INTERMITTENT CATHETER Back to Search Results
Model Number 277716
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Codes Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 08/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the last lot of products customers received have the diameter bigger than the other lot they got.It caused pain in insertion and bleeding when using them.No medical intervention was reported.
 
Additional Manufacturer Narrative
The reported event was unconfirmed as the device met specifications.No root cause was provided as the reported event was not reportable.As the reported event was unconfirmed, a device history record review was not required.As the reported event was unconfirmed, labelling review was not required.Correction: d, g, h h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the last lot of products customers received was item 277716.The diameter on those was bigger than the other lot they got.It caused pain insertion and bleeding when using them.No medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® RUBBER UTILITY CATHETER
Common Device Name
INTERMITTENT CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15362735
Report Number1018233-2022-06917
Device Sequence Number12138707
Product Code KOD
UDI-Device Identifier00801741034114
UDI-Public(01)00801741034114
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2026
Device Model Number277716
Device Catalogue Number277716
Device Lot NumberNGFW2970
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/25/2022
Supplement Date Received by Manufacturer02/08/2023
Initial Report FDA Received Date09/06/2022
Supplement Report FDA Received Date02/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-