Model Number 9735542
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information regarding a thermal therapy system being used in a soft tissue ablation (neuro) procedure.It was reported that there was a software error during the t map session.The t-maps came over and they were swapped.They were swapped for the second t-map session.It happened 3-4 times.There was no reported delay to the procedure.No reported impact to the patient.The manufacturer representative believes they set the tmap sequence up to run the sagittal image first and the coronal image second.The first ablation went fine, they stopped the scan and disconnected from the scanner.They copied and pasted the same sequence and reran it for the next ablation.The images were swapped (cor #1 and sag #2), the rep had to use the swapped images and swapped target buttons.The images kept swapping with each new sequence (not every time, but almost every time).When post processing the movies there was one log file that was swapping the damage.
 
Additional Manufacturer Narrative
Other relevant device(s) are: product id: m450001b022, version #: 3.3.2, the manufacturer representative went to the site to test the navigation system.No hardware parts were replaced.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
H3: a software investigation analysis was initiated to determine the probable cause of the issue through log analysis.Analysis found that the reported issue was not cause by the software.Analysis found that the software functioned as designed.Codes b01, c19, d14 are applicable to this analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
H3: a software analysis was initiated to determine the probable cause of the issue through log analysis.Analysis found that the software was functioning as designed.However, the reported event was related to a software issue at the digital intercommunication of medicine (dicom) information level, and the issue was confirmed to originate in the mri scanner system of le ilt v29.1 systems.This issue was documented in a medtronic navigation software anomaly tracking database.H6: fdm b01, fdr c10, and fdc d02 are applicable to the software analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISUALASE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15362766
Report Number1723170-2022-01424
Device Sequence Number12053284
Product Code GEX
UDI-Device Identifier00643169792890
UDI-Public00643169792890
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K081656
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9735542
Device Catalogue Number9735542
Device Lot NumberN07831809
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2022
Supplement Date Received by Manufacturer11/22/2022
02/10/2023
Initial Report FDA Received Date09/06/2022
Supplement Report FDA Received Date11/29/2022
02/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/11/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10...."
Patient Age26 YR
Patient SexFemale
Patient Weight48 KG
-
-