Model Number MI1000 MED-EL CONCERT PIN
Medical Device Problem Codes Impedance Problem (2950); Output Problem (3005)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Event or Problem Description
The user's family perceived in april 2022 that the user was not presenting auditory perception with the device.No trauma has been reported.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Additional Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However, to determine an exact root cause a device investigation of the explanted device is necessary.
 
Event or Problem Description
The users family perceived on (b)(6) 2022 that the user was suddenly not presenting auditory perception with the device.Reimplantation is considered, however, no date has been scheduled yet.
 
Event or Problem Description
The user_s family perceived in april 2022 that the user was suddenly not presenting auditory perception with the device.No trauma has been reported.Reimplantation is considered, however, no date has been scheduled yet.
 
Additional Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However, to determine an exact root cause a device investigation of the explanted device is necessary.The device was explanted but has not been received for investigational purposes yet.
 
Additional Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However, to determine an exact root cause a device investigation of the explanted device is necessary.The device was explanted but has not been received for investigational purposes yet.
 
Event or Problem Description
The user_s family perceived in april 2022 that the user was suddenly not presenting auditory perception with the device.No trauma has been reported.Correction: the user has been reimplanted.The concerned device has not been recieved for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key15362781
Report Number9710014-2022-00651
Device Sequence Number5165154
Product Code MCM
UDI-Device Identifier09008737083625
UDI-Public(01)09008737083625
Combination Product (Y/N)N
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/04/2024
Device Model NumberMI1000 MED-EL CONCERT PIN
Device Catalogue Number08841
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2022
Supplement Date Received by Manufacturer09/05/2022
09/05/2022
09/05/2022
Initial Report FDA Received Date09/06/2022
Supplement Report FDA Received Date11/17/2022
10/02/2023
10/02/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age28 MO
Patient SexMale
-
-