• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HOL L 10MM ENDO APPLIER; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HOL L 10MM ENDO APPLIER; APPLIER, SURGICAL, CLIP Back to Search Results
Model Number IPN915197
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.
 
Event or Problem Description
The jaws of the applier were found misaligned upon opening the kit.The user tested the applier, but it was difficult to load a clip and the clip did not easily leave the applier.Therefore, a new unit was used instead.
 
Event or Problem Description
The jaws of the applier were found misaligned upon opening the kit.The user tested the applier, but it was difficult to load a clip and the clip did not easily leave the applier.Therefore, a new unit was used instead.
 
Additional Manufacturer Narrative
(b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was originally produced at the manufacturing facility as part of a 50 pc.Lot in december of 2021.Evaluation of the returned instrument as received shows that the tips of the jaws are aligned in the open and closed position.Functional testing of this instrument was performed, and it is able to pick-up, retain, close and release multiple clips both with and without the use of silastic test tubing.There were no non-conforming features found on this device.We are unable to validate this complaint.All 50 instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOL L 10MM ENDO APPLIER
Common Device Name
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15362879
Report Number3011137372-2022-00179
Device Sequence Number5124650
Product Code GDO
UDI-Device Identifier24026704696837
UDI-Public24026704696837
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN915197
Device Catalogue Number544995
Device Lot Number06J2140658
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2022
Supplement Date Received by Manufacturer10/04/2022
Initial Report FDA Received Date09/06/2022
Supplement Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/14/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-