Implanted date: device was not implanted.Implanted date: device was not implanted.Initial reporter occupation: occupation: lead tech.The photo of actual product provided by the user was confirmed.One catheter fracture piece and the elongated coil were found.The actual samples upon receipt were three fracture pieces (sample a, b and c).The following investigation was performed for each sample.Sample a: catheter fractured piece.Sample b: catheter fractured piece and coil fractured piece.Sample c: coil fractured piece.Based on the investigation results, and since sample b was not found in the photo provided, it was possible that the actual sample was artificially cut after the photo was taken.[sample a].Visual inspection found that the rear end had been fractured and the hub side was not returned.It was found that the total length was approximately 650mm.Magnifying inspection found that the reinforcement had been exposed at the fractured section.No anomaly such as a crush was found in other sections.Then, the fractured section was confirmed from the top surface.It was found that the fractured section had an oval shape, and a torn shape was found.Electron microscopic inspection found no scratch on the side surface, the fracture surface was found to be smooth.X-ray fluoroscopic inspection of the lumen did not find any anomaly including an obstruction.[sample b].Visual inspection found that it was the catheter fracture piece with elongated coil.The following investigation was performed with end "a" (where the coil protruded) and end "b" (the other end).It was found that the total length was approximately 1215mm.Th catheter part was approximately 75mm and the wire part was approximately 1140mm.From the investigation results, it was found that the total length of sample a (length: approximately 650mm) and the catheter fracture piece of sample b (length: approximately 75mm) was found to be approximately 725mm.Since the normal product is approximately 1000mm, it was inferred that the actual catheter was missing approximately 275mm.Magnifying inspection found that the reinforcement had been exposed at end "a".It had been crushed at approximately 25mm and 35mm from end "a".The reinforcement had been exposed at end "b" the fractured section "a" was confirmed from the top surface.It was found that the fractured section "a" had an oval shape, and a torn shape was found.The fractured section "b" was confirmed from the top surface.It was found that the fractured section "b" had an oval shape.Electron microscopic inspection found a circumferential scratch was found in the vicinity of end "a".No scratch was found in other sections.The fracture surfaces of end "a" and "b" were both smooth.From the investigation results, it was found that the fracture surface of sample a and the fracture surface of end "a" of sample b were similar.Therefore, it was likely that the fractured section found in the photo of actual product provided by the user was end "b" of sample b.X-ray fluoroscopic inspection of the lumen found that the coil had been jumbled and had been got stuck at approximately 17mm from end "a".The coil was removed from the fractured piece and magnifying inspection of the distal end of coil was performed.It was found that the red foreign substance had been adhered.Qualitative analysis was performed by ft-ir for the red foreign substance and found in the investigation results.It was confirmed that it was very similar to the peak of a representative protein.Due to this, it was highly possible that the red foreign substance was blood.Ft-ir (fourier transform infrared spectroscopy method): an analysis method to confirm the molecular structure of an object by analyzing the spectrum obtained by irradiating the measurement target with infrared rays.Since the wavelength of the infrared absorbed is almost unique to each substance, qualitative analysis of the substance is possible by comparing it with the infrared standard absorption spectrum.The inner and outer diameters of the catheter were measured and confirmed to meet the specifications.[sample c].Visual inspection found that it was a fractured piece of elongated coil.It was found that the total length was approximately 890mm.Review of the manufacturing record and the shipping inspection record of the involved product code/lot number combination confirmed that there were not any indications of anomaly in them.A search of the complaint file found no other similar report with the involved product code/lot number combination from other facilities.Based on the investigation result, no anomaly was found in the manufacturing record and the shipping inspection record.Therefore, the following causes of occurrence were inferred for getting stuck, crushing, and fracturing of the actual sample, respectively.It was likely that it got stuck due to jumbling of the coil and adhesion of blood clots.However, it was not possible to identify the cause of jumbling in the coil.After the packaging process of this product, 100% visual inspection is performed, and the crush found in the actual sample could be detected by visual inspection.Therefore, it was inferred that some external force was applied to the actual sample after the product was opened.It was highly likely that it was cut at the involved section after the actual sample was removed.It was inferred that it was caused by the application of clipping force while some hard and sharp object was in contact with the actual sample.It was highly likely that it was cut at the involved section after the actual sample was removed.The actual catheter was likely to have a rear end side missing by approximately 275mm.However, it was unlikely that the rear end side would remain in the patient's body.Ashitaka factory is always continuing diligence in maintaining the product quality by performing following work and inspections.In the tube molding process, the core wire of which the outer diameter is strictly controlled is covered with resin, and then the wire is braided over the resin.Subsequently, the core wire is removed from the braided tube.By this production method, the inner diameter of the catheter can be firmly assured.In the packaging and cartoning processes, dedicated tools and containers are used for handling this product to protect the product and to prevent the occurrence of the crush.Before the packaging process, 100% visual inspection of the catheter is performed to assure that there is no anomaly including a crush.Terumo medical products (tmp) (importer) registration no.(b)(4) is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.(b)(4).
|