MAUDE Adverse Event Report: TELEFLEX MEDICAL PE ADULT-PED DRY/ WET LF 6/CS; BOTTLE, COLLECTION, VACUUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL PE ADULT-PED DRY/ WET LF 6/CS; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number IPN028442
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
Customer reported that blood is leaking out of the quick disconnect.
 
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.
 
Additional Manufacturer Narrative
Qn#: (b)(4).The device history record has been reviewed and no issues or discrepancies were found which could potentially relate to this complai nt.No nonconformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled & inspected according to our specifications.One photo of catalog number a-6000-08lf (pe adult-ped dry/ wet lf 6/cs) was received for analysis.It was observed the pleur evac unit with two arrows, one pointing to the suction port of ats connector and the other arrow pointing between the red and blue connector of ats connector.Also, it was observed a blue medical pad under ats connector, and such pad has a red stain that it seems blood, that give the impression that the ats connector is leaking.The photo is not enough clear to observe a leak on ats connector.No corrective action can be established at this moment since the product sample is not available for evaluation.Product sample is necessary to perform a proper investigation to determinate the root cause and the corresponding corrective actions.Customer complaint cannot be confirmed based only on the photo provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved on this complaint.
 
Event or Problem Description
Customer reported that blood is leaking out of the quick disconnect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PE ADULT-PED DRY/ WET LF 6/CS
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15363187
Report Number3004365956-2022-00068
Device Sequence Number17763928
Product Code KDQ
UDI-Device Identifier34026704631231
UDI-Public34026704631231
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/24/2025
Device Model NumberIPN028442
Device Catalogue NumberA-6000-08LF
Device Lot Number74B2201462
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2022
Supplement Date Received by Manufacturer09/27/2022
Initial Report FDA Received Date09/06/2022
Supplement Report FDA Received Date10/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-