| Model Number |
5943 |
| Medical Device Problem Codes |
Loss of or Failure to Bond (1068); Crack (1135)
|
| Health Effect - Clinical Code |
Fall (1848)
|
| Date of Event |
09/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during surgery the carbon from the metal end of the table top came apart causing the patient to fall.One of the doctors caught the patient's head and there were currently no known injuries observed.There were no treatments or extended stay needed for the patient.Point of contract stated there was a 1' crack on the carbon, and they removed it from service.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery the carbon from the metal end of the table top came apart causing the patient to fall.One of the doctors caught the patient's head and there were currently no known injuries observed.There were no treatments or extended stay needed for the patient.Point of contract stated there was a 1' crack on the carbon, and they removed it from service.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the limited available information and per the analysis of internal historical data pertaining to the product, it was found that there were no preventative maintenance checks performed from manufacturer's end.Internal records show that only one table was delivered to this facility in the year 2004 and hence, we suspect that the spine top joints may have disbonded due to aging and corrosion.
|
| |
|