• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7015
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.A 3.0mm x 15mm quantum maverick balloon catheter was selected for used.However, upon unpacking, the device was fractured.The procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The device was visually and microscopically examined.There were numerous hypotube kinks to the device.At 56.7cm and 60.9cm from the strain relief there were two separations of the hypotube.The separated ends were jagged and ovaled in shape indicating the device was kinked prior to separation.There was contrast in the inflation lumen and there was blood in the guidewire lumen.The balloon was tightly folded, and blood was present within the balloon folds.Product analysis confirmed the reported break as there were two separations to the hypotube of the device.
 
Event or Problem Description
It was reported that shaft break occurred.A 3.0mm x 15mm quantum maverick balloon catheter was selected for used.However, upon unpacking, the device was fractured.The procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15365794
Report Number2124215-2022-32233
Device Sequence Number17711466
Product Code LOX
UDI-Device Identifier08714729392781
UDI-Public08714729392781
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2024
Device Model Number7015
Device Catalogue Number7015
Device Lot Number0027077056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/19/2022
Supplement Date Received by Manufacturer09/15/2022
Initial Report FDA Received Date09/06/2022
Supplement Report FDA Received Date10/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexUnknown
Patient Weight55 KG
-
-