| Model Number |
20E |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that the device is booting up to a white display.A white display can be indicative of a device lock up condition.As a result, defibrillation therapy may be unavailable, if it was needed.There was no patient involvement reported with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that the device is booting up to a white display.A white display can be indicative of a device lock up condition.As a result, defibrillation therapy may be unavailable, if it was needed.There was no patient involvement reported with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was retuned to stryker for evaluation and was unable to duplicate the reported issue.The device's software was reloaded and the ui stack flex assembly was replaced as a precaution.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
|
| |
|