| Model Number |
640HX0204 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The healthcare professional reported that during an endovascular embolization procedure, the physician tried to deploy an orbit galaxy helical xtrasoft coil (640hx0204 / 30723400) through a 150cm x 5cm, 45-degree prowler select plus microcatheter (606s155fx / 30763314), but while advancing the coil midway through, the physician encountered resistance inside the microcatheter.The physician was unable to push the coil forward and was unable to retract the coil out from the microcatheter.As a result, the physician removed the microcatheter and the coil together.The procedure was completed using a new microcatheter and new coil.On 24-aug-2022, additional information was received.The information indicated that the target vessel was the anterior communicating artery (acom).The complaint devices did not appear damaged at any time.Nothing was noted to be obstructing the microcatheter.A continuous flush was maintained through the microcatheter.Based on the additional event information on 24-aug-2022, the event has been deemed reportable as a ¿malfunction.¿.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Procode is krd/hcg.The initial reporter first name was not reported.The initial reporter phone: (b)(6).The initial reporter email address was not reported.The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.A review of manufacturing documentation associated with this lot (30723400) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.This is one of 2 products involved with the reported complaint.The associated manufacturer report numbers are: 3008114965-2022-00567.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that the product was received in the product analysis lab on 13-sep-2022.The returned product is pending evaluation.A supplemental 3500a report will be submitted once the product investigation has been completed.Updated sections: b.4, d.9, g.3, g.6.H.2, h.3, h.6, and h.10.This is one of 2 products involved with the reported complaint.The associated manufacturer report numbers are: 3008114965-2022-00566 and 3008114965-2022-00567.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report the investigational finding of the returned device.[conclusion]: the healthcare professional reported that during an endovascular embolization procedure, the physician tried to deploy a 2mm x 4cm orbit galaxy helical xtrasoft coil (640hx0204 / 30723400) through a 150cm x 5cm, 45-degree prowler select plus microcatheter (606s155fx / 30763314), but while advancing the coil midway through, the physician encountered resistance inside the microcatheter.The physician was unable to push the coil forward and was unable to retract the coil out from the microcatheter.As a result, the physician removed the microcatheter and the coil together.The procedure was completed using a new microcatheter and new coil.On 24-aug-2022, additional information was received.The information indicated that the target vessel was the anterior communicating artery (acom).The complaint devices did not appear damaged at any time.Nothing was noted to be obstructing the microcatheter.A continuous flush was maintained through the microcatheter.The complaint device was returned for evaluation and analysis.The investigation finding is documented below.Investigation summary: a non-sterile 2mm x 4cm orbit galaxy helical xtrasoft coil was received contained in the decontamination pouch.Visual inspection was performed.There was no appearance of damage observed.Microscopic inspection was performed.Under magnification, the embolic coil was observed not attached to the device.This detached condition was not originally reported in the complaint, and the exact time the detachment occurred cannot be determined.There was no elongation marks observed that suggest the coil was mechanically detached.The concomitant 45-degree prowler select plus microcatheter (606s155fx / 30763314) was flushed until water was observed coming out from the distal end.Then the returned 2mm x 4cm orbit galaxy helical xtrasoft was introduced into the microcatheter and advanced.The 2mm x 4cm orbit galaxy helical xtrasoft was advanced until it came out from the distal tip of the microcatheter without any noticeable resistance.The functional test was performed with the concomitant 45-degree prowler select plus microcatheter.The returned 2mm x 4cm orbit galaxy helical xtrasoft was able to pass through the entire length of the microcatheter without significant resistance.Additionally, no other damages were found on the orbit galaxy that could have contributed to the issue reported during the procedure.The reported issue documented in the complaint cannot be confirmed with the evidence available.It is possible that other clinical and procedural factors that cannot be replicated during the analysis may have contributed to the reported failure.A review of manufacturing documentation associated with this lot (30723400) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) does contain the following recommendations: never advance, withdraw, or torque the delivery tube against resistance without first determining the cause of resistance under fluoroscopy.Manipulation of the delivery tube against resistance can cause damage and/or premature detachment of the coil.If unusual friction is noted within the infusion catheter, remove the detachable coil system.If friction is noted with any subsequent detachable coil system, carefully examine the detachable coil system and the infusion catheter for possible damage.Replace both if necessary.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Updated sections: b.4, g.3, g.6.H.2, h.3, h.6, and h.10.This is one of 2 products involved with the reported complaint.The associated manufacturer report numbers are: 3008114965-2022-00566 and 3008114965-2022-00567.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|
Search Alerts/Recalls
|
|