| Model Number |
MI1210 SYNCHRONY ST |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user reported that he could hardly hear with the device.An x-ray will be taken.Re-implantation is considered.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
The user reported that he could hardly hear with the device.The user was re-implanted.
|
| |
|