Catalog Number 328447
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/12/2022
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Initial reporter name and address: the customer's address is unknown.(b)(6) has been used as a default.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd safetyglide¿ insulin syringe was missing the thumb press tab.This complaint was created to capture the 2nd of 2 related incidents.The following information was provided by the initial reporter: we have a decent number of insulin syringes in h20 that are missing the tab at the end of the plunger to help draw it back.
 
Event Description
It was reported that the bd safetyglide¿ insulin syringe was missing the thumb press tab.This complaint was created to capture the 2nd of 2 related incidents.The following information was provided by the initial reporter: "we have a decent number of insulin syringes in h20 that are.Missing the tab at the end of the plunger to help draw it back.".
 
Manufacturer Narrative
H6: investigation summary: customer returned an image of 0.5ml, 31 guage, 6mm sg syringe from lot 1343773, cat # 328447.The image shows the plunger rod is missing its thumbpress.The damage at the base of the plunger rod suggests it may have been cut or broken off.A review of the device history record was completed for batch# 1343773.All inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.Bd was able to confirm the customer¿s indicated failure.A definitive root cause cannot be determined.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SAFETYGLIDE¿ INSULIN SYRINGE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15368403
MDR Text Key301939052
Report Number1920898-2022-00606
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K992734
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number328447
Device Lot Number1343773
Was Device Available for Evaluation? No
Date Manufacturer Received09/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/09/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-