• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA REVERSE SHOULDER SYSTEM THREADED GLENOID BASEPLATE Ø24.5X25; SHOULDER GLENOID BASEPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA REVERSE SHOULDER SYSTEM THREADED GLENOID BASEPLATE Ø24.5X25; SHOULDER GLENOID BASEPLATE Back to Search Results
Model Number 04.01.0190
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery performed 2 months after the primary due to threaded glenoid baseplate mobilization.The surgeon explanted the baseplate, screws, glenosphere, pe liner and humeral metaphysis.
 
Additional Manufacturer Narrative
Batch review performed on 02-sep-2022.Lot 2113957: (b)(4) items manufactured and released on 09-feb-2022.Expiration date: 2027-01-30.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event during the period of review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVERSE SHOULDER SYSTEM THREADED GLENOID BASEPLATE Ø24.5X25
Common Device Name
SHOULDER GLENOID BASEPLATE
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15368480
Report Number3005180920-2022-00681
Device Sequence Number15022201
Product Code PHX
UDI-Device Identifier07630040708098
UDI-Public07630040708098
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K171058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.01.0190
Device Catalogue Number04.01.0190
Device Lot Number2113957
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2022
Initial Report FDA Received Date09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-