Model Number 100/199/065
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample and five pictures were received for evaluation.The sample was received in used condition with its original packaging open.Visual and functional testing were performed.Visual inspection found no discrepancies were observed.Functional testing found a leak was observed from the cuff, the complaint was confirmed.The most probable root cause is that the cuff was damaged after it left the manufacture facility due to the product is 100% leak tested.This remediation mdr was generated under protocol (b)(4), as a result of warning letter (b)(4).
 
Event or Problem Description
It was reported that the nurse found a cuff leak during testing.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ENDOTRACHEAL TUBE W/HOLDER AND CONNECTOR
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15368924
Report Number3012307300-2022-18053
Device Sequence Number6323646
Product Code BTR
UDI-Device Identifier15019315022319
UDI-Public15019315022319
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K790312
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100/199/065
Device Catalogue Number100/199/065
Device Lot Number3845493
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Initial Report FDA Received Date09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-