MAUDE Adverse Event Report: SYNTHES GMBH ADJUSTABLE PARALLEL WIRE GUIDE; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH ADJUSTABLE PARALLEL WIRE GUIDE; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Model Number 312.73
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Part # 312.730.Synthes lot # h627346.Supplier lot # h627346.Release to warehouse date: 10 oct 2018.Supplier: avalign technologies-nemcomed.No ncrs were generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from australia reports an event as follows: it was reported that on (b)(6) 2022, the adjustable guide has broken away from the main parallel guide body.There was no reported patient interaction.No further information is available.This report is for an adjustable parallel wire guide.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6: photo investigation: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned photo.Visual analysis of the photo revealed that parallel-guide f/guide-wires ø1.6 adjust had the tightening nut broken.The breakage condition was observed at the shaft of the guide sleeve.All parts of the device were observed in the evidence provided.H3, h6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that parallel-guide f/guide-wires ø1.6 adjust was observed with the tightening nut broken.The breakage condition was observed at the shaft of the guide sleeve.All parts of the device were received in the evidence provided.A dimensional inspection for the parallel-guide f/guide-wires ø1.6 adjust was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the parallel-guide f/guide-wires ø1.6 adjust would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h6: a device history record (dhr) review was conducted: part # 312.730, synthes lot # h627346, supplier lot # h627346, release to warehouse date: 10 oct 2018, supplier: (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADJUSTABLE PARALLEL WIRE GUIDE
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 
3035526892
MDR Report Key15370383
Report Number8030965-2022-06480
Device Sequence Number11171634
Product Code FZX
UDI-Device Identifier10886982187901
UDI-Public(01)10886982187901
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number312.73
Device Catalogue Number312.730
Device Lot NumberH627346
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/12/2022
Supplement Date Received by Manufacturer09/28/2022
Initial Report FDA Received Date09/07/2022
Supplement Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-