| Model Number |
1260 |
| Medical Device Problem Codes |
Fracture (1260); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Cardiac Arrest (1762)
|
| Date of Event |
08/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that cardiac arrest and distal tip damaged occurred.A mach1 diagnostic guide catheter was selected for use in a coronary intervention procedure.The distal tip was crushed damaged and a blood clot has occurred, the blood flow was temporarily interrupted, resulting in cardiac arrest requiring hospital treatment.No further patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter address: (b)(6).Initial reporter city: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to h6 device code.Initially reported as a fracture.Supplement mdr is correcting h6 code to material integrity problem.Initial reporter address 1: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that cardiac arrest and distal tip damage occurred.A mach1 diagnostic guide catheter was selected for use in a coronary intervention procedure.The distal tip was crushed damaged and a blood clot has occurred, the blood flow was temporarily interrupted, resulting in cardiac arrest requiring hospital treatment.No further patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the model-6f mach 1 q3.5 sh (side holes) was returned for analysis.Visual and microscopic observation was performed on the catheter hub, shaft, and distal tip.A kink was identified approximately 6.5 cm from the distal tip and the distal tip was identified damaged/crushed.No further damaged were observed.Correction to h6 device code.Initially reported as fracture.Supplement mdr is correcting h6 code to material integrity problem.Initial reporter address 1: (b)(6); e1 initial reporter city: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that cardiac arrest and distal tip damaged occurred.A mach1 diagnostic guide catheter was selected for use in a coronary intervention procedure.The distal tip was crushed damaged and a blood clot has occurred, the blood flow was temporarily interrupted, resulting in cardiac arrest requiring hospital treatment.No further patient complications were reported.
|
| |
|