Model Number V200DNUS
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device did not detect breathing.No patient injury was reported.
 
Additional Manufacturer Narrative
A product sample was received for evaluation.Visual and functional testing were performed.Upon visual inspection, it was observed that there was one screw cover missing.During function testing where the device was hooked up to an o2 supply and turned on, the reported problem was duplicated.The root cause of the reported issue was found to be a faulty demand valve and detector block.These faults are due to normal wear and tear.Repair actions included replacing the faulty components, recalibration, and cleaning of device.After repairs, device passed functional tests.A device history record (dhr) review was conducted which indicated previous service was for preventative maintenance.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC PARAPAC VENTILATOR
Common Device Name
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK  LU3 4BU
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK   LU3 4BU
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15371020
Report Number3012307300-2022-18135
Device Sequence Number6090105
Product Code BTL
UDI-Device Identifier35019315074183
UDI-Public35019315074183
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K030803
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberV200DNUS
Device Catalogue Number130003V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2022
Initial Report FDA Received Date09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-