• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL FAST-CATH¿ HEMOSTASIS INTRODUCER, CATH-LOCK¿ LOCKING HUB, 12 CM SHEATH, 6 F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL FAST-CATH¿ HEMOSTASIS INTRODUCER, CATH-LOCK¿ LOCKING HUB, 12 CM SHEATH, 6 F; INTRODUCER, CATHETER Back to Search Results
Model Number 406701
Medical Device Problem Codes Fracture (1260); Material Frayed (1262)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
This report is to advise of a introducer fracture found during analysis.
 
Additional Manufacturer Narrative
One 6f fast-cath introducer sheath was received for evaluation.The sheath tubing had been cut and also stretched and torn into three sections; all three sections were returned.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the sheath tubing damage remains unknown.
 
Additional Manufacturer Narrative
The physician's name is dr.(b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH¿ HEMOSTASIS INTRODUCER, CATH-LOCK¿ LOCKING HUB, 12 CM SHEATH, 6 F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15373375
Report Number3005334138-2022-00505
Device Sequence Number19231998
Product Code DYB
UDI-Device Identifier05414734216944
UDI-Public05414734216944
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K061015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model Number406701
Device Catalogue Number406701
Device Lot Number7555533
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer09/11/2022
Initial Report FDA Received Date09/07/2022
Supplement Report FDA Received Date09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-